Product Description
The BiPAP Synchrony provides more natural and comfortable noninvasive ventilation specifically for COPD patients and others suffering from respiratory insufficiency. ALS or neuromuscular patients may also benefit from the therapy provided by Synchrony.
Synchrony can be adapted to meet a patient's needs by offering flexibility in treatment options. The unit can be upgraded from Spontaneous Mode (S) to a version with Spontaneous/Timed Mode (S/T). For greater flexibility, the unit has continuous positive airway pressure, Timed- and Pressure-Control Modes.
Digital Auto-Trak Sensitivity™ has enhanced leak compensation to allow for better matching of the patient's breathing pattern. Vent Ramp and RiseTime™ also aid in the delivery of comfortable ventilation. Features a broad pressure range of 4 to 30 cm H2O and the ability to collect patient therapy, compliance and machine data using an optional built-in modem.
Traveling is easy with the Synchrony, as it is DC capable and weighs only six pounds.
Options
- BiPAP Synchrony S without modem 1002816
- BiPAP Synchrony S with modem 1002817
- BiPAP Synchrony S/T without modem 1003986
- BiPAP Synchrony S/T with modem 1003987
Product Highlights
- Digital Auto-Trak Sensitivity
- Ensures optimum sensitivity despite changing breathing patterns and leak conditions.
- Pressure-based leak determination provides a highly responsive and accurate triggering and cycling performance.
- Automatically detects mouth leaks during IPAP, and effectively cycles to EPAP.
- Automatically adjusts sensitivity levels and increases patient comfort.
- Easy-To-Use intuitive interface makes setup and operation quick and easy; displays key set and measured parameters.
- Compact, Lightweight design weighs only six pounds and uses minimal space on nightstand.
- Adjustable RiseTime enhances patient comfort and allows for adjustment of inspiratory flow rate to meet patient needs.
- Upgrade To Timed Modes minimizes investment costs until COPD patient demonstrates need for Timed Modes of support. System design provides flexibility for current and future purchase decisions.
- Wide Pressure Range and Flow Capability provides maximum pressure delivery of 30 cm H2O and higher flow capabilities for use with a wider range of patients.
- Vent Ramp feature designed specifically for the bi-level ventilation user to allow unit pressure to gradually ramp to prescription pressure with each breath.
- Optional/Integrated Alarms provide audible and visual alarms for High/Low Pressure, Power Loss, Low Battery, Optional Apnea, Patient Disconnect and Minute Volume.
- Oxygen Valve provides safe operation with supplemental oxygen by turning the oxygen flow off/on with unit operation to prevent excessive build-up of oxygen when the unit is off.
- Monitoring Of Data, such as Tidal Volume and Minute Ventilation, provides patients and caregivers with the knowledge of proper treatment and ventilator set up.
- Optional Built-In Modem allows transfer data from a remote location.
- Encore® Software™ stores months of patient usage and machine data for download and reporting.
- DC-Power Compatibility allows for easy connection to an external battery. Alarm also notifies user when battery is low.
- Universal Power Supply simplifies travel worldwide.
Specifications
- Control Accuracy
- IPAP 4 to 30 H2O ± 2 cm H2O
- EPAP 4 to 25 H2O ± 2 cm H2O
- CPAP 4 to 20 H2O ± 2 cm H2O
- Breath Rate 0 to 30 BPM (PC and S/T) 4 to 30 BPM (T) Greater of 1 BPM or +/- 10% of setting when measured over a four minute period.
- Timed Inspiration 0.5 to 3.0 sec ± (0.10% of setting + 0.1 sec)
- Rise Time 100 to 600 msec ± 25%
- Vent Ramp Disabled, 0.5, 1.0, 2.0, 3.0 cm H2O per breath (0.1+ 0.25 of setting) cm H2O/breath. Measured over the full duration of the programmed ramp.
- Physical
- Dimensions 12" L x 7" W x 6" H
- Weight 6 lbs
- Electrical
- AC Voltage 100 to 240 V, 50/60 Hz
- DC Voltage 12 V
- AC Current 1.25 A Maximum
- DC Current 5.5 A Maximum
- Trigger and Cycle
- Spontaneous Trigger
- Shape Trigger
- Volume 6 cc Above V Leak
- Spontaneous Cycle
- Spontaneous Expiratory Threshold (SET)
- Shape Cycle
- IPAP Maximum of 3.0 sec
- Flow Reversal
- Mode of Operation
- BiPAP Spontaneous, or Spontaneous/Timed, Timed, Pressure Control, and CPAP